Description
Description
MAX-12 Maxtec Replacement O2 Sensor for Siemens, code R109P45
REPLACEMENT FOR
- Analytical Industries[PSR-11-917-M]
- City Technology[MOX-9]
- Datascope[0600-00-0002 (MAX-12f)]
- EnviteC[OOM-202]
- Hamilton[396008, 396009, Galileo, Raphael, Arabella, Aladdin, 51254, G5, Arabella]
- IMT[PF300]
- International Technologies[M-04]
- MSA[472062, 470644]
- Mercury Medical[10-103-06]
- Pacifitech[PT-12A]
- Sensidyne[Acutonic Hi Freqency Jet Ventilator, 00M102]
- Siemens[9004979-E347E]
- Spacelabs[90513, Capnograph]
- Taema[YR049700, Cesar-2, Horus-4]
- Teledyne[R24MED #B-62321]
- Teleflex (Hudson)[5556, V-12]
- VTI[103100]
SPECS
- Range of Measurement:0 to 100%
- Output Range:12.5 to 16.5 mV
- Zero Offset:≤ 0.250 mV
- Response Time:≤ 12 s
- Linearity:≤ 3%
- Expected Lifetime:> 1,000,000 %O2 hours
- Temperature Compensation:NTC
- Operational Temperature Range:10Ëš to 40ËšC
- Storage Temperature Range:-20Ëš to 50ËšC
- Humidity Range:0% to 90% RH, non-condensing
- Ambient Pressure Range:700-1250 mbar
- Interference:PER EN ISO 80601-2-55
- Electrical Interface:3-pin Molex
Reviews